GC-MS Analysis Tool

Retention Index Calculator

Calculate Kováts retention indices for plant volatile organic compounds (PVOCs) using n-alkane standards. Automatically match calculated RIs against a curated library of 505 essential oil compounds.

Van den Dool & Kratz formula for temperature-programmed GC
Built-in RI library with 505 plant volatile compounds
Real-time linear regression validation (R² analysis)
Gas Chromatography Illustration

Retention Index Formula

Van den Dool & Kratz (1963) — Temperature-Programmed GC
RI = 100 × [n + (tR'(unknown) − tR'(n)) / (tR'(n+1) − tR'(n))]
tR'(unknown)

Adjusted retention time of unknown compound (RT − void time)

n

Carbon number of the n-alkane eluting just before the unknown

tR'(n)

Adjusted RT of n-alkane with n carbons

tR'(n+1)

Adjusted RT of n-alkane with n+1 carbons

Workflow Overview

1
Enter Void Time
Dead time t₀ from unretained peak
2
Input Alkanes
RTs for C₇–C₂₅ standards
3
Add Samples
Unknown compound RTs
4
View Results
Calculated RI values
5
Match Library
505 compound database

Column Void Time (t₀)

Enter the void time (dead time) of unretained analytes on your GC column.

= 1.58 min

Instructions: Enter retention times for n-alkane standards (C₇-C₂₅). You may enter values in minutes or seconds. Enter at least 5 consecutive alkanes for reliable calibration. Verify R² value > 0.99 before analyzing samples.

Alkane Carbon # (n) RT (minutes) RT (seconds) Typical RT (min) Status
Calibration Curve (RT vs Carbon Number)
Calibration Statistics
Standards Entered: 0
R² Value: —
Slope: —
Intercept: —
Awaiting data entry

Sample Retention Times

Enter retention times for unknown compounds. Enter up to 30 compounds.

# RT (minutes) RT (seconds) Adjusted RT (sec) Compound Name (optional)

Note: Enter retention times in either minutes or seconds. The other column will be calculated automatically. Adjusted RT = RT - void time. Make sure alkane standards are entered and R² > 0.99 before proceeding to Results.

Calculated Retention Indices & Library Matches

(for compound matching)
# RT (min) Adj RT (sec) Calculated RI Library RI Δ RI Library Compound Plant Source Reference

Enter alkane standards and sample data to view results

Go to tab 2 (Alkane Standards) to begin

RI Library Browser

Quick RI range:
505 compounds in library
RI Compound Plant Source Reference
Plant Source Codes:
CEGR - Ceanothus greggii
CELAS2 - Ceratoides lanata
CHPU4 - Chrysothamnus pulchellus
DAFO - Dalea formosa
DYAC - Dyssodia acerosa
KOSP - Koeberlinia spinosa
FLCE - Flourensia cernua
PSSC6 - Psorothamnus scoparius
ANCA - Artemisia species
GUSA - Gutierrezia sarothrae

How to Use This Calculator

1
Determine Column Void Time

Inject an unretained compound (e.g., methane, butane for solvent delay) or use the solvent front peak to determine t₀. Enter this value on the Alkane Standards tab.

2
Run n-Alkane Standard Mix

Separate a standard mixture of n-alkanes (C₇-C₂₅ or similar range covering your compounds) using the exact same GC method you will use for samples.

3
Enter Alkane Retention Times

Enter the retention times for each alkane in either minutes or seconds. Check the calibration curve and verify that R² > 0.99. R² values below 0.99 indicate chromatographic problems.

4
Enter Sample Data

Enter retention times for unknown peaks from your sample chromatogram. You may enter times in either minutes or seconds.

5
View Results & Match Compounds

Go to the Results tab to see calculated RIs and nearest matches from the library. A Δ RI ≤ 5 units combined with a good mass spectral match (≥ 950) typically indicates positive identification.

Important Notes
  • • Compounds with RIs < 850 or > 1900 may show greater variability. A tolerance of ± 10 units may be appropriate for early- or late-eluting compounds.
  • • This calculator uses the van den Dool & Kratz formula for temperature-programmed GC. For isothermal separations, use the original Kováts formula based on adjusted retention volumes.
  • • Retention indices are column-specific. Library values were determined on DB-5 (5% phenyl methyl siloxane) columns.
  • • Always confirm identifications with mass spectral matching and/or authentic standards when possible.
Reference

Lucero, M., Estell, R., Tellez, M., & Fredrickson, E. (2009). A Retention Index Calculator Simplifies Identification of Plant Volatile Organic Compounds. Phytochemical Analysis, 20(5), 378-384. DOI: 10.1002/pca.1137